Editor-in-chief
Maria Stella Graziani

Deputy Director
Martina Zaninotto

Associate Editors
Ferruccio Ceriotti
Davide Giavarina
Bruna Lo Sasso
Giampaolo Merlini
Martina Montagnana
Andrea Mosca
Paola Pezzati
Rossella Tomaiuolo
Matteo Vidali

International Advisory Board Khosrow Adeli Canada
Sergio Bernardini Italy
Marcello Ciaccio Italy
Eleftherios Diamandis Canada
Philippe Gillery France
Kjell Grankvist Sweden
Hans Jacobs The Netherlands
Eric Kilpatrick UK
Magdalena Krintus Poland
Giuseppe Lippi Italy
☩Howard Morris Australia
Mario Plebani Italy
Sverre Sandberg Norway
Ana-Maria Simundic Croatia
☩Jill Tate Australia
Tommaso Trenti Italy
Cas Weykamp The Netherlands
Maria Willrich USA
Paul Yip Canada


Publisher
Biomedia srl
Via L. Temolo 4, 20126 Milano

Responsible Editor
Giuseppe Agosta

Editorial Secretary
Arianna Lucini Paioni
Biomedia srl
Via L. Temolo 4, 20126 Milano
Tel. 0245498282
email: biochimica.clinica@sibioc.it



Area soci
Non possiedi o non ricordi la password!
Clicca qui

BC: Articoli scritti da R. Paleari

Il progetto pilota SIBioC di VEQ della misura dell’emoglobina glicata
Pilot SIBioC EQAS project on glycated hemoglobin measurement
<p>Two fresh blood samples collected with EDTA&nbsp;were distributed by courier in December 2014 to 206 Italian laboratories asking for the determination of their HbA<sub>1c</sub> concentrations. Target HbA<sub>1c</sub> values were assigned by the IFCC reference measurement procedure based on HPLC capillary&nbsp;electrophoresis. The results, collected from 193 laboratories using analytical systems mainly from five&nbsp;manufacturers (Bio-Rad Laboratories, A. Menarini Diagnostics, Roche Diagnostics, Sebia and Tosoh), showed a&nbsp;global variability (in terms of CV) of 5.3% and 3.8% at HbA<sub>1c</sub> values of 37.4 mmol/mol (sample 1) and 62.0 mmol/mol&nbsp;(sample 2), respectively. Globally, 84% of the participants reported HbA<sub>1c</sub> results within the total allowable error (TE)&nbsp;of 8.6% (sample 1) and 93% for sample 2. These percentages decreased to 70% and 77%, respectively, when using&nbsp;a goal for the allowable TE of 6.0% as criterion. Inter-laboratory CVs, calculated per group of methods, were between&nbsp;3.3% and 5.0% and between 2.2% and 3.7% for sample 1 and 2, respectively. Tosoh users registered the smaller&nbsp;inter-laboratory CV in sample 1 and Sebia&rsquo;s in sample 2. With regard to trueness, all methods had a mean bias &le;2.8%&nbsp;respect to the target values, with the exception of Tosoh (bias of +6.1% and +5.8%, for samples 1 and 2, respectively).</p>
Biochimica Clinica ; 39(6) 568-574
Contributi scientifici - Scientific Papers
 
Valutazione del sistema Capillarys 2 Flex Piercing per la misura dell’emoglobina A1c
Evaluation of the Capillarys 2 Flex Piercing system for the determination of hemoglobin A1c (HbA1c)
R. Paleari  |  A. Mosca  | 
<p>The&nbsp;Capillarys HbA<sub>1c</sub> kit implemented on the Capillarys 2 Flex Piercing platform uses capillary electrophoresis to separate&nbsp;and quantify HbA<sub>1c</sub> in human blood. We performed an evaluation of this system by checking imprecision, relative bias&nbsp;and robustness respect to the analysis of samples with variable total hemoglobin concentrations. Intra-assay CVs were&nbsp;between 1.1% and 2.8% for HbA<sub>1c</sub> values between 35.8 mmol/mol and 96.4 mmol/mol. Inter-assay CVs, evaluated on&nbsp;control materials, were 2.0% and 2.4 % for high (68.6 mmol/mol) and low (36.8 mmol/mol) control levels, respectively.&nbsp;Results in three blood samples with various concentrations of HbA<sub>1c</sub> (36, 60 and 87 mmol/mol) were not affected by&nbsp;variation in total hemoglobin concentrations (between 40 to 180 g/L). Only at very low total hemoglobin concentration,&nbsp;the imprecision was slightly higher (CV 3.1%). The results obtained by capillary electrophoresis (y-method) were well&nbsp;correlated with those obtained by the HPLC Tosoh G8 (x-method) (y = 0.73 + 0.978x, r=0.998, n=100).</p>
Biochimica Clinica ; 38(2) 110-114
Contributi scientifici - Scientific papers
 
Valutazione multicentrica dell’analizzatore Tosoh G8 per la misura dell’emoglobina A2 e dell’emoglobina F
Multicenter evaluation of the Tosoh G8 analyzer for determination of hemoglobin (Hb) A2 and F
<p>The analytical performance of the new Tosoh automated analyzer HLC-723 G8 (-thalassemia analysis mode) to determine hemoglobin variants and to measure HbA2 and HbF in human blood was evaluated in three Italian centres. The within- and between-run imprecision for HbA2 were good, with CV between 0.2% and 1.8% and between 0.9% and 5.4%, respectively. The CV for HbF was between 0.4% and 9.8% (within-run) and beetwen 0.8% and 13.1% (between-run). The comparability of HbA2 measurements between different centres was excellent (r=0.99), but a significant bias in comparison with the previous version of the instrument was noted. Experiments to test the HbA2 stability in blood confirmed that blood samples stored at -80 &deg;C were stable for at least 4 months and that storage at -20 &deg;C is not recommended. In conclusion, the Tosoh G8 analyser was found reliable and robust and, therefore, suitable for the measurement of HbA2 and HbF in human blood.</p>
Biochimica Clinica ; 37(1) 30-35
Contributi Scientifici - Scientific Papers
 
Determinazione della variabilità biologica dell'emoglobina A2
Determination of biological variation of hemoglobin A2.
R. Paleari  |  M. Montagnana  |  E. Danese  |  M. Tozzi  |  G.C. Guidi  |  A. Mosca  | 
<p><strong>Determination of biological variation of hemoglobin A2.</strong> We present an experimental report aimed to evaluate the biological variation of hemoglobin A2 (HbA2), a minor hemoglobin component in post-natal life, accounting for 2.5%-3.5% of the total hemoglobin in red cells, which is very relevant for the laboratory diagnosis of thalassemic syndromes. We took five blood specimens from 17 apparently healthy subjects (9 men and 8 women, ages 26-52 years) on the same day, every two weeks for two months. Samples were stored at -80 &#176;C until analysis and assayed in duplicate by Bio-Rad Variant II analyzer. Data were analyzed by the ANOVA. There were no differences in HbA2 values between men and women. HbA2 exhibited marked individuality: within- (CVI) and between-subject (CVG) biological variation were 0.7% and 7.7%, respectively. Desirable analytical goals derived from biological variation for imprecision (0.5 CVI), bias [0.25 (CVI2 + CVG2)1/2] and total error [1.65 (0.5 CVI) + 0.25 (CVI2 + CVG2)1/2]were 0.4%, 1.9%, and 3.1%, respectively. In conclusion, this is the first evidence that HbA2, as well as total hemoglobin, is under a strict homeostatic control. Our data also show that stringent analytical goals are needed for the clinical application of HbA2 measurements.</p>